Contact Us For Quotation | Email: 464560351@qq.com | Whatsapp&Wechat:+8618620373879 |
Main Features
• Microprocessor controlled high voltage insulation resistance tester with Diagnostic functions. |
Specifications
Insulation resistance (*): Accuracy is not guaranteed with setting of 250V or less
Rated voltage | 500V | 1000V | 2500V | 5000V | 10000V | 12000V |
---|---|---|---|---|---|---|
Max measurement value | 500GΩ | 1TΩ | 2.5TΩ | 5TΩ | 35TΩ | |
Accuracy | 400k~50GΩ ±5%rdg±3dgt | 800k~100GΩ 5%rdg±3dgt | 2M~250GΩ ±5%rdg±3dgt | 4M~500GΩ ±5%rdg±3dgt | 8M~1TΩ ±5%rdg±3dgt | |
50.1G~500GΩ ±20%rdg (*) | 101G~1TΩ ±20%rdg | 251G~2.5TΩ ±20%rdg | 501G~5TΩ ±20%rdg | 1.01T~10TΩ ±20%rdg | ||
10.1T - 35TΩ Values are displayed, but accuracy isn't guaranteed | ||||||
Short circuit current | Max 5.0mA | |||||
Load resistor to output rated voltage | 0.5MΩ or more | 1MΩ or more | 2.5MΩ or more | 5MΩ or more | 20MΩ or more | 24MΩ or more |
Output voltage
Rated voltage | 500V | 1000V | 2500V | 5000V | 10000V | 12000V |
---|---|---|---|---|---|---|
Monitor accuracy | ±10%±20V | ±10%±20V | ±10%±20V | ±10%±20V | ±10%±20V | ±10%±20V |
Output accuracy | 0 - +20% | 0 - +10% | 0 - +10% | 0 - +10% | -5 - +5% | -5 - +5% |
Selectable range | 50 - 600V (in steps of 5V) | 610 - 1200V (in steps of 10V) | 1225 - 3000V (in steps of 25V) | 3050 - 6000V (in steps of 50V) | 6100 - 10000V (in steps of 100V) | 10100 - 12000V (in steps of 100V) |
Voltage measurement
Measuring range | DCV:±30 ~ ±600V、ACV:30 ~ 600V (50/60Hz) |
---|---|
Accuracy | ±2%rdg±3dgt |
Current measurement
Measuring range | 5.00nA - 2.40mA (Depending on the insulation resistance) |
---|
Capacitance measurement
Measuring range | 5.0nF ~ 50.0μF (500V/1000V/2500V/5000V) | 5.0nF - 1.0µF (display range: 5.0nF - 50.0µF) (10000V/12000V) |
---|---|---|
Accuracy | ±5%rdg±5dgt |
General
Applicable standards IEC 61010-1 CAT IV 600V Pollution degree 2,
IEC 61010-031, IEC 61326, IEC 60529 (IP64): with the lid closed.Operating temperature & humidity range -10ºC - Relative humidity 85% or less
50ºC /(when operating with an external power supply, no condensation)
0ºC - 40ºC /Relative humidity 85% or less
(when operating with battery, no condensation)Storage temperature & humidity range -20ºC - 60ºC / Relative humidity 75% or less (no condensation) Power source Rechargeable Lead storage battery (12V)*
Charging time: approx. 8 hours / AC Power supply (100V - 240V, 50/60Hz)
* Continuous measuring time:
approx. 4 hours
(a load of 100MΩ at the Insulation resistance 12000V Range.)
approx. 2 hours
(Max current consumption in range while rated currents are maintained.)Dimensions 330 (L) × 410 (W) × 180 (D) mm *Instrument and Hard case Weight 9kg approx. (including battery) *Instrument and Hard case Included Accessories 7170 (Main cord), 7224A (Earth cord), 7225A (Guard cord), 7226A (Line probe),
7227A (Line probe with alligator clip), 8029 (Extension prod),
8212-USB-W (USB adaptor with KEW Windows (Software)),
Instruction manualOptional Accessories 7254 (Longer Line probe with alligator clip)